Particulate | |
Measuring range | 0 … 1000 µg/m³ |
Accuracy | PM2.5 @ -10…+60 °C ± 10 µg/m³ @ 0 … 100 µg/m³ ± 10 % of measuring @ 100 … 1000 µg/m³ PM10 @ -10…+60 °C ± 25 µg/m³ @ 0 … 100 µg/m³ ± 25 % of measuring @ 100 … 1000 µg/m³ (with no fog) |
Temperature | |
Measuring range | -40 … +60 °C |
Accuracy | ± 0.1 °C @ -40 ... +60 °C @ moved air > 2 m/s |
Rel. Humidity | |
Measuring range | 0 … 100 % rel. Humidity |
Accuracy | Typ. ± 1,5 % r. H. @ 25 °C and < 80 % r. H. 2 % r. H. over the complete measuring range |
Air pressure | |
Measuring range | 300 … 1200 hPa |
Accuracy | ± 0.25 hPa @ -20 … +60°C @ 800 … 1100 hPa ± 0.50 hPa @ -20 … +60°C @ 600 … 800 hPa |
Interface | |
Type | RS 485 HD |
Data format | 8N1, 8N2, 8E1, 8E2, 8O1, 8O2, 7E1, 7N1 |
Baud rate | 2400 ... 57600 Baud |
Operating voltage | |
Battery connection | 2x 12 ... 30 V DC |
General | |
Ambient conditions | -20 ... +70 °C 0 ... 100 rel. humidity, including condensation |
Current load | Measurement electronics: typ. 360 mW max. 500 mW Ventilation: 2 W |
Housing | Stainless steel, polycarbonate |
Protection | IP53 |
Dimension | Ø 120 x 317 |
Weight | aprox. 1.25 kg |
Data output digital | |
Electr. output | THIES ASCII |
General | |
Electr. connection | Measuring electronics with plug, weather and radiation protection with permanently connected 5 m cable. |
Data output digital | |
Electr. output | MODBUS RTU |
General | |
Electr. connection | Measuring electronics with plug, weather and radiation protection with permanently connected 5 m cable. |